Thermocouple Temperature Sensor – Types K, J, N, E, T to EN 60584

A thermocouple temperature sensor measures temperature from the voltage generated at the junction of two dissimilar metals. Thermocouples withstand high temperatures and vibration and respond quickly – ideal for furnaces, kilns, dryers and flue gas. ISE supplies type E, J, K, N and T thermocouples in simplex, duplex and triplex versions, including ATEX/IECEx models – to the European standard EN 60584-1.

Industrial thermocouple temperature sensor

Thermocouple temperature sensor types E, J, K, N, T

Thermocouple temperature sensor EMF for types E, J, K, N, T
Thermocouple temperature sensor EMF for types E, J, K, N, T
Type Class 2 tolerance range (EN 60584-1) Applications
K (NiCr–NiAl) −40 … 1,200 °C Most common; furnaces, flue gas
N (NiCrSi–NiSi) −40 … 1,200 °C More stable than K at high temperature
J (Fe–CuNi) −40 … 750 °C Reducing atmospheres, plastics
E (NiCr–CuNi) −40 … 900 °C Highest EMF
T (Cu–CuNi) −40 … 350 °C Low temperature, food, refrigeration

Thermocouple temperature sensor specifications

Parameter Value
Sensor type Type E · J · K · N · T
Configuration Simplex · Duplex · Triplex
Wiring 2-wire + matching extension cable (EN 60584-3)
Accuracy class Class 1 · Class 2 per EN 60584-1
Approvals ATEX Ex d II 2 GD · IECEx Ex d · UKEX Ex d · ATEX Ex ia II 1 GD · IECEx Ex ia · UKEX Ex ia
Connection head Aluminium · 316 stainless · Plastic
Termination Terminal block · Temperature transmitter · Tails
Indicator Battery powered · Loop powered
Thread size 1/4″ · 1/2″ · 3/4″ · 1″ · M20 × 1.5
Thread type BSPP · BSPT · NPT · Metric
Sheath material 304, 310, 316 stainless · Alloy · Inconel · Ceramic

Thermocouple tolerance per EN 60584-1

Type K and Pt100 temperature sensor tolerance per EN 60584-1 and EN 60751
Type K and Pt100 temperature sensor tolerance per EN 60584-1 and EN 60751

Type K class 1: ±1.5 °C or ±0.004·|t| (−40 … 1,000 °C); class 2: ±2.5 °C or ±0.0075·|t| (−40 … 1,200 °C), whichever is greater. Thermocouples use classes 1 and 2 – classes A, B and 1/3 DIN apply to Pt100 only.

Wiring and cold-junction compensation

Pt100 2, 3 and 4-wire connection and type K extension cable
Pt100 2, 3 and 4-wire connection and type K extension cable

Use the matching extension cable (type K: green per EN 60584-3) with correct polarity; the instrument must provide cold-junction compensation. For long runs use a 4–20 mA temperature transmitter.

Standards and approvals

Standard Scope
EN 60584-1 Thermocouple EMF tables and tolerances
EN 60584-3 Extension cables and colour code
EN 61515 Mineral-insulated thermocouples
ATEX 2014/34/EU, IECEx Equipment for explosive atmospheres
EN IEC 60079-1 / -11 Ex d / Ex ia

Manufacturers

EXCEL INSTRUMENTS – India · S&S INSTRUMENTS – India · HEINZ – Germany · H&B SENSOR – United Kingdom.

Related products

Contact ISE

EQUIPMENT INDUSTRIAL SOLUTION CO., LTD (ISE) – Ho Chi Minh City, Vietnam.

Contact us: Sales +84 909 617 185 – Technical +84 979 117 185 – Email: [email protected]