Pt100 Temperature Sensor (RTD) to EN 60751

A Pt100 temperature sensor (RTD) uses a platinum element with 100 Ω at 0 °C, giving high accuracy and stability across −196 … +650 °C. ISE supplies Pt100 and Pt1000 sensors in simplex, duplex and triplex versions; 2, 3 and 4-wire; class A, B and 1/3, 1/5, 1/10 DIN; including ATEX/IECEx explosion-proof versions – to the European standard EN 60751.

Pt100 RTD sensor RTD320

How a Pt100 temperature sensor works

Pt100 RTD temperature sensor resistance curve per EN 60751
Pt100 RTD temperature sensor resistance curve per EN 60751

Platinum resistance rises almost linearly with temperature, about 0.385 Ω/°C: 100 Ω at 0 °C and 138.51 Ω at 100 °C, defined in EN 60751 by the Callendar–Van Dusen equation.

Pt100 temperature sensor specifications

Parameter Value
Sensor type Pt100 · Pt1000
Configuration Simplex · Duplex · Triplex
Wiring 2-wire · 3-wire · 4-wire
Accuracy class Class B · Class A · 1/3 DIN · 1/5 DIN · 1/10 DIN
Approvals ATEX Ex d II 2 GD · IECEx Ex d · UKEX Ex d · ATEX Ex ia II 1 GD · IECEx Ex ia · UKEX Ex ia
Connection head Aluminium · 316 stainless · Plastic
Termination Terminal block · Temperature transmitter · Tails
Indicator Battery powered · Loop powered
Thread size 1/4″ · 1/2″ · 3/4″ · 1″ · M20 × 1.5
Thread type BSPP · BSPT · NPT · Metric
Range Low (−196 … +100 °C) · Standard (−50 … +450 °C) · High (−50 … +650 °C)

Pt100 accuracy classes per EN 60751

Type K and Pt100 temperature sensor tolerance per EN 60584-1 and EN 60751
Type K and Pt100 temperature sensor tolerance per EN 60584-1 and EN 60751
Class Tolerance (°C) At 0 °C At 100 °C
Class A ±(0.15 + 0.002·|t|) ±0.15 °C ±0.35 °C
Class B ±(0.3 + 0.005·|t|) ±0.3 °C ±0.8 °C
1/3 DIN ⅓ of class B ±0.1 °C ±0.27 °C
1/10 DIN ⅒ of class B ±0.03 °C ±0.08 °C

Source: JUMO – Pt100 accuracy class (classes A, B); 1/3 and 1/10 DIN are fractions of class B over narrower ranges.

2, 3 and 4-wire Pt100 connection

Pt100 2, 3 and 4-wire connection and type K extension cable
Pt100 2, 3 and 4-wire connection and type K extension cable

3-wire is the industrial standard; for long cable runs use 4-wire or a head-mounted 4–20 mA temperature transmitter.

Standards and approvals

Standard Scope
EN 60751 Industrial platinum resistance thermometers, tolerance classes
ATEX 2014/34/EU, IECEx Equipment for explosive atmospheres
EN IEC 60079-1 / -11 Flameproof Ex d / intrinsic safety Ex ia
DIN 43772 Thermowells